TracerCo radiation survey meters are new to Canada, and represent an extremely robust, reasonably priced instrument for multiple purposes. The new NORM Meter (T-407-C) offers detection of Alpha and Beta particles as well as Gamma radiation.
Business Update article on the TracerCo NORM-IS monitor
This flexible system pairs a thin window Geiger-Müller probe for external contamination surveys, with a 1" Sodium Iodide scintillator offering great sensitivity for NORM detection through vessel walls.
The T-407-C is intrinsically safe to CSA standards. It is also weather resistant, preventing the penetration of rain, through most surfaces.
The T-407-C also allows display in Bq/cm2 for NORM radionuclides Lead-210 and Radium-226, allowing direct comparison to the Canadian Guidelines for the Management of Naturally Occurring Radioactive Material.
Official User Manual |
Radiation Detected |
Scintillator: gamma, high energy beta GM: Alpha, beta with some gamma response |
Measurement Modes |
Scintillator: CPS, uSv/hr GM: CPS, Bq/cm2 All modes have background subtration option |
Dose Rate Range (Scintillator Probe) | 0.000 to 50uSv/hr |
Count Range |
Scintillator: 0-1,000,000 CPM GM: 0.00 to 240,000 CPM |
OverRange Response |
Bar graph display will read full scale. Digital numeric display will read "over" |
Integrate Period |
Auto = 60 seconds or 1000 counts User Defined = 5-600 seconds |
Scintillator Detector | NaI crystal in metal/polymer enclosure |
GM Detector | Single halogen thin window detector in static dissipative nylon housing |
Handset Material | Static Dissipative Nylon |
Weight | Handset: 500g, Scintillator: 700g Gm: 435g |
Battery | Alkaline Maganese 9V |
Battery Life |
Scintillator: 85 hours typical GM: 190 hours typical |
Low battery indication | <10 hours available life remaining |
Variation with battery voltage | +/- 2% |
Working temperature range | -20 to +50 degress celcius |
Variation with temperature | <10% |
Humidity range | 0 - 95% |
Ingress protection rating |
Scintillator: IP67 GM: IP34 Handset IP65 |
Hazardous area certification code |
Class 1, GPS A,B,C & D T4 Class 1, Zone 0, Ex ia IIC T4 Class 1, Zone 0 AEx ia IIC T4 |
Kit contains holster, for hands free use while climbing | ![]() |