TracerCo T-407-C NORM Meter

 

TracerCo T-407-C NORM Meter Alpha Beta Gamma

TracerCo radiation survey meters are new to Canada, and represent an extremely robust, reasonably priced instrument for multiple purposes. The new NORM Meter (T-407-C) offers detection of Alpha and Beta particles as well as Gamma radiation. 

Business Update article on the TracerCo NORM-IS monitor

This flexible system pairs a thin window Geiger-Müller probe for external contamination surveys, with a 1" Sodium Iodide scintillator offering great sensitivity for NORM detection through vessel walls.

The T-407-C is intrinsically safe to CSA standards. It is also weather resistant, preventing the penetration of rain, through most surfaces.

The T-407-C also allows display in Bq/cm2 for NORM radionuclides Lead-210 and Radium-226, allowing direct comparison to the Canadian Guidelines for the Management of Naturally Occurring Radioactive Material.

 

Official User Manual

 

Radiation Detected

Scintillator: gamma, high energy beta

GM: Alpha, beta with some gamma response

Measurement Modes

Scintillator: CPS, uSv/hr

GM: CPS, Bq/cm2

All modes have background subtration option

Dose Rate Range (Scintillator Probe) 0.000 to 50uSv/hr
Count Range

Scintillator: 0-1,000,000 CPM

GM: 0.00 to 240,000 CPM
OverRange Response

Bar graph display will read full scale.

Digital numeric display will read "over"

Integrate Period

Auto = 60 seconds or 1000 counts

User Defined = 5-600 seconds

Scintillator Detector NaI crystal in metal/polymer enclosure
GM Detector Single halogen thin window detector in static dissipative nylon housing
Handset Material Static Dissipative Nylon
Weight Handset: 500g, Scintillator: 700g Gm: 435g
Battery  Alkaline Maganese 9V
Battery Life

 Scintillator: 85 hours typical

GM: 190 hours typical

Low battery indication <10 hours available life remaining
Variation with battery voltage +/- 2%
Working temperature range -20 to +50 degress celcius
Variation with temperature <10%
Humidity range 0 - 95%
Ingress protection rating

Scintillator: IP67

GM: IP34

Handset IP65

Hazardous area certification code

Class 1, GPS A,B,C & D T4

Class 1, Zone 0, Ex ia IIC T4

Class 1, Zone 0 AEx ia IIC T4

Kit contains holster, for hands free use while climbing  tracerCo In Holster

 


 

 