This gamma probe is excellent for anything requiring a high sensitivity measurement of energies from 60KeV to 2 MeV. The detector of choice for environmental monitoring
Detector
1" Sodium iodide Scintillator
GENERAL | Intended for general gamma (dose) measurements |
---|---|
Compatibility | All WB Johnson Instruments (others on request) |
Operating Voltage | 600-900 vdc |
Wall Thickness | 40 mg/cm2 |
Range | 0-50 uSv/hr |
CPM -uR/hr (Cs-137) | 175 |
Energy Response | 60keV - 2 MeV gamma |
Sensitive Area | 1" x 1" |
Finish | Black Polyurethane |
Dimensions | 1-7/8" dia x 6" L |
Environmental | -20° F[-28° C]-140° F[60° C] | 5-95% RH IPX4 Housing |
Weight | 8 oz |